A computer model has been developed to evaluate the accuracy of an oscillometric method to measure the arterial pulse pressure from a patient with a rotary ventricular assist device (VAD). This computer model consists of three major components: the cardiovascular system, the HeartMate II VAD, and the operation of an automated cuff. Simulation was performed to mimic failure, recovery, and normal cardiac functions of a patient, supported by the HeartMate II VAD at different levels from minimum to maximum. The oscillating cuff pressure, simulating the air pressure of a deflecting cuff, was obtained from simulation under different conditions to test the accuracy of an oscillometric algorithm in determining the arterial pulse pressure. The algorithm was able to detect the systolic and diastolic arterial pressure with the error within ±2 mmHg in most cases, except the cases when ventricular suction, induced by the VAD, occurred. The results from this study suggested that the oscillometric algorithm is capable to accurately detect the arterial pulse pressure for a rotary VAD patient if the algorithm is properly tuned.
机译:已经开发了一种计算机模型来评估示波法的准确性,该示波法可通过旋转心室辅助设备(VAD)测量来自患者的动脉脉压。该计算机模型包括三个主要部分:心血管系统,HeartMate II VAD和自动套囊的操作。由HeartMate II VAD在最小到最大的不同水平上模拟患者的衰竭,恢复和正常心脏功能。通过在不同条件下的仿真获得模拟袖套气压的摆动袖带压力,以测试示波法确定动脉脉压的准确性。该算法能够检测出大多数情况下的收缩压和舒张动脉压,误差在±2 mmHg之内,除了由VAD引起的心室吸引的情况外。这项研究的结果表明,如果对算法进行了适当的调整,示波法能够准确地检测出旋转VAD患者的动脉搏动压力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号