System for application of contrast therapy and Transcutaneous Electrical Nervous Stimulation for pain treatment

机译:对比疗法和经皮神经电刺激治疗疼痛的系统

获取原文

摘要

There are two methods widely used by physiotherapists for pain treatment and rehabilitation purposes. These methods are: Thermal contrast therapy and Transcutaneous Electrical Nervous Stimulation (TENS). Thermal contrast therapy usually consists on the application of hot and cold pads over the site of interest during a determined period of time (usually less than 30 minutes). Sometimes, the hot and cold sessions are intercalated for better therapeutic results. The problem with the application of such therapy nowadays is that it is based on pads which must be heated or cooled by mean of external devices (ovens, hot bath and freezers) using energy during the whole day to keep the pads at the correct temperature ready for their use. Another problem is the loss of thermal energy of the pads during their use on the patient skin. Thus, if at the beginning of the therapy the cold (or hot) pad was at the right temperature, after few minutes the temperature will not be enough for the therapeutic result. Furthermore, usually these thermal therapies are applied simultaneously with TENS, thus the physiotherapist requires at least two devices, making their application cumbersome. This paper presents an electronic system for the application of heat or cold and TENS therapies simultaneously. Design and preliminary results on the characterization of the system are presented showing the viability of the proposal. Temperatures over the range of 8 °C - 50 °C were reached and maintained independently of the ambient temperature. TENS was generated with a voltage range of 10 - 50V and frequency from 1 - 300Hz and current output until 50 mA.
机译:物理治疗师广泛使用两种方法来进行疼痛治疗和康复。这些方法是:热对比疗法和经皮神经电刺激(TENS)。热对比疗法通常包括在确定的时间段内(通常少于30分钟)在目标部位上应用热垫和冷垫。有时,可插入冷热疗程以获得更好的治疗效果。如今,这种疗法的问题在于其基于垫子,该垫子必须在一整天内使用外部设备(烤箱,热水浴和冷冻箱)进行加热或冷却,以保持垫子处于正确的温度下供他们使用。另一个问题是垫在患者皮肤上使用期间的热能损失。因此,如果在治疗开始时冷(或热)垫处于正确的温度,则几分钟后温度将不足以达到治疗效果。此外,通常这些热疗法与TENS同时应用,因此物理治疗师需要至少两个装置,这使得它们的应用繁琐。本文介绍了一种同时应用冷热和TENS疗法的电子系统。提出了系统表征的设计和初步结果,显示了该建议的可行性。达到并维持在8°C-50°C范围内的温度与环境温度无关。 TENS的产生电压范围为10-50V,频率为1-300Hz,电流输出直至50 mA。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号