An EoN (Electrical impedance spectroscopy-On-a-Needle) with bipolar interdigitated electrodes at the very end of a hypodermic needle has been introduced. The device can be utilized to analyze the electrical impedance of biotissues for the purpose of discriminating between tissues prior to biopsy. A flexible photomask and photoresist spray coating were employed in the lithography process to realize the electrode pattern on the round surface of the needle. To evaluate the discrimination capability of the EoN, the impedance of the leg fat and muscle from a slaughtered chicken was measured over the frequency range 10 kHz to 1 MHz. The resistance and reactance were classified into two groups, which was in good agreement with the reported permittivity of fat and muscle tissues from bovines.
机译:引入了在皮下注射针头末端带有双极叉指电极的EoN(针式电阻抗谱)。该装置可用于分析生物组织的电阻抗,以在活检之前区分组织。在光刻过程中采用了柔性光掩模和光致抗蚀剂喷涂,以在针的圆形表面上实现电极图案。为了评估EoN的辨别能力,在10 kHz到1 MHz的频率范围内测量了屠宰鸡腿脂肪和肌肉的阻抗。电阻和电抗分为两类,这与所报道的牛脂肪和肌肉组织的介电常数非常吻合。