MRI scanner is the most important diagnostic tool that aids in the diagnosis of diseases and offers most sensitive and non-invasive way of Imaging. The sophisticated tool helps clinician to understand the cause of patient's particular health issue. RF Surface coils are used in MRI for better signal reception as they are more sensitive to signal close to the coil, thus detecting signals from organ of interest more efficiently. Hence attempts are made in order to improve the performance of the coil with respect to its signal to noise ratio, penetration depth, Specific Absorption Rate, etc. In this paper, we have designed, simulated and fabricated two designs of MRI surface coil for 1.5 Tesla scanner. A balun is designed to feed the unbalanced coaxial signal to a balanced coil. Significant improvement in the Q factor and reduction of the bandwidth was observed in Design B which improves SNR of the coil as compared to traditional surface coil Design A. Simulation and experimental result have been compared to validate the proposed design.
机译:MRI扫描仪是最重要的诊断工具,可帮助诊断疾病,并提供最灵敏,无创的成像方式。先进的工具可帮助临床医生了解患者特殊健康问题的原因。射频表面线圈在MRI中用于更好的信号接收,因为它们对靠近线圈的信号更敏感,从而更有效地检测到来自感兴趣器官的信号。因此,为了提高线圈在信噪比,穿透深度,比吸收率等方面的性能,进行了一些尝试。在本文中,我们设计,模拟和制造了两种用于MRI表面线圈的设计,分别适用于1.5特斯拉扫描仪。平衡-不平衡转换器设计用于将不平衡同轴信号馈入平衡线圈。与传统的表面线圈设计A相比,设计B中的Q因子显着提高,带宽减小,与传统的表面线圈设计A相比,线圈的SNR有所提高。已对仿真和实验结果进行了比较,以验证所提出的设计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号