Inter-leg Distance Measurement as a Tool for Accurate Step Counting in Patients with Multiple Sclerosis

机译:腿间距离测量作为多发性硬化症患者准确步骤计数的工具

获取原文

摘要

Step detection is commonly performed using wearable inertial devices. However, methods based on the extraction of signals features may deteriorate their accuracy when applied to very slow walkers with abnormal gait patterns. The aim of this study is to test and validate an innovative step counter method (DiSC) based on the direct measurement of inter-leg distance. Data were recorded using an innovative wearable system which integrates a magneto-inertial unit and multiple distance sensors (DSs) attached to the shank. The method allowed for the detection of both left and right steps using a single device and was validated on thirteen people affected by multiple sclerosis (0 < EDSS < 6.5) while performing a six-minute walking test. Two different measurement ranges for the distance sensor were tested (DS200: 0–200 mm; DS400: 0–400 mm). Accuracy was evaluated by comparing the estimates of the DiSC method against video recordings used as gold standard. Preliminary results showed a good accuracy in detecting steps with half the errors in detecting the step of the instrumented side compared to the non-instrumented (mean absolute percentage error 2.4% vs 4.8% for DS200; mean absolute percentage error 2% vs 5.4% for DS400). When averaging errors across patients, over and under estimation errors were compensated, and very high accuracy was achieved (E%<1.2% for DS200; E%<0.7% for DS400). DS400 is the suggested configuration for patients walking with a large base of support.
机译:步骤检测通常使用可穿戴惯性装置进行。然而,基于信号特征的提取的方法可能会使它们在具有异常步态图案的慢速步行者施加到非常慢的步行者时劣化。本研究的目的是基于腿间距离的直接测量来测试和验证创新的步骤计数器方法(盘)。使用创新的可穿戴系统记录数据,该系统集成了磁性惯性单元和附接到柄部的多个距离传感器(DSS)。允许用于检测使用单个设备左侧和右侧的步骤的方法,和对受多发性硬化症(0 200 :0-200毫米; DS. 400 :0-400毫米)。通过将光盘方法的估计与用作金标准的视频录制进行比较来评估精度。初步结果在与无仪表(平均绝对百分比误差2.4%VS 4.8%的DS的平均绝对率误差2.4%Vs 4.8%)检测仪表侧的步骤中的误差中的步骤检测步骤的初步精度 200 ;平均直观百分比误差2%与DS的5.4% 400 )。当患者平均误差时,补偿估计误差并在估计误差下进行补偿,并且实现了非常高的准确性(例如DS <1.2% 200 ; E.DS <0.7% 400 )。 DS. 400 是用大量支持的患者行走的建议配置。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号