首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Feasibility Study of a Syringe Angulation Measurement System Suitable for Ophthalmic Regional Anesthesia Training
【24h】

Feasibility Study of a Syringe Angulation Measurement System Suitable for Ophthalmic Regional Anesthesia Training

机译:注射器角度测量系统适用于眼科局部麻醉培训的可行性研究

获取原文

摘要

Regional anaesthesia for the eye requires a needle that is inserted in a specific way into the eye orbit. Angulation of the needle is a concern for the ophthalmic anaesthetists. Inappropriate angulation can place the needle in an unsafe location or result in the unsafe release of the anaesthetic. Recently developed mannequin based ophthalmic training systems had a visual angulation warning; however, an automated quantifiable warning and the ability to provide a preemptive warning to the trainee would provide useful knowledge. In this paper, an angulation measurement methodology which uses two external cameras, in the mannequin system, which can accurately sense angulation in both the vertical and the horizontal axis of the syringe with customizable angle warning system, is presented. Details of the system developed, test results and the outcome are discussed.
机译:眼睛的区域麻醉需要针对特定​​方式插入眼眶轨道的针。针的角度是眼科麻醉师的关注。不恰当的角度可以将针放在不安全的位置或导致麻醉剂的不安全释放。最近开发的基于时装模特的眼科训练系统具有视觉角度警告;然而,自动化的可量化警告和向受训人员提供先发制人的警告的能力将提供有用的知识。在本文中,呈现了使用两个外部相机的角度测量方法,其可以提出了可以在具有可定制角度警告系统的垂直和水平轴的垂直和水平轴上准确地感测角度的角度测量方法。讨论了系统的细节,测试结果和结果。