In this paper we present a method for the estimation of leaks in non-invasive ventilation. Accurate estimation of leaks is a key component of a ventilator, since it determines the ventilator performance in terms of patient-ventilator synchrony and air volume delivery. In particular, in non-invasive ventilation, the patient flow is significantly different from the flow measured at the ventilator outlet. This is mostly due to the vent orifice along the tube that is used for exhalation, but also to the non-intentional leaks that occur elsewhere in the circuit (e.g., at the mask). Such leaks are traditionally quantified via a model with two parameters, but only one of them is continually updated - the other is fixed. The new algorithm allows for breath-by-breath update of both parameters. This was made possible by leveraging a model describing the patient respiratory mechanics.
机译:在本文中,我们提出了一种估计非侵入性通气中泄漏的方法。精确估计泄漏是呼吸机的关键部件,因为它决定了患者通风机同步和空气输送方面的呼吸机性能。特别地,在非侵入性通气中,患者流程与在呼吸机出口处测量的流动显着不同。这主要是由于沿着用于呼气的管的通气孔,而且还是在电路(例如,在掩模处的其他地方发生的非有意的泄漏。这种泄漏传统上通过具有两个参数的模型来量化,但是它们中只有一个不断更新 - 另一个是固定的。新算法允许逐呼吸的两个参数更新。通过利用描述患者呼吸力学的模型来实现这一点。