首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Hemodynamic Sensing of 3D Fingertip Force using PPG device on Proximal Part
【24h】

Hemodynamic Sensing of 3D Fingertip Force using PPG device on Proximal Part

机译:PPG装置在近端部分使用PPG装置的血流动力学感应

获取原文

摘要

This research proposed a novel method to estimate the 3D contact force of a fingertip using a photoplethysmogram (PPG) device on the proximal part of a finger. The proposed method detects non-pulsatile and pulsatile components of PPG signals, extracts eight feature values related to the contact force, and estimates the force by multiple linear regression. In the validation experiments, the participants wore a PPG device at the proximal part of the index finger and applied contact force. Then, the relationship between the contact force and PPG feature values was investigated. As a result, the system was found to be capable of estimating the 3D contact force with a root-mean-square error of 3.5% of the maximum contact force.
机译:该研究提出了一种新的方法来估计指尖的3D接触力在手指的近端部分上使用光电电敏感图(PPG)装置的倾斜纤维装置。所提出的方法检测PPG信号的非脉冲和脉动分量,提取与接触力有关的八个特征值,并通过多个线性回归估计力。在验证实验中,参与者在食指的近端部分佩戴PPG装置和应用接触力。然后,研究了接触力和PPG特征值之间的关系。结果,发现该系统能够估计具有3.5%的最大接触力的根均方误差的3D接触力。