首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Free Induction Decay MR Signal Measurements toward Ultra-low Field MRI with an Optically Pumped Atomic Magnetometer
【24h】

Free Induction Decay MR Signal Measurements toward Ultra-low Field MRI with an Optically Pumped Atomic Magnetometer

机译:使用光学泵浦原子磁力计自由诱导衰减MR信号测量对超低场MRI

获取原文
获取外文期刊封面目录资料

摘要

Ultra-low field magnetic resonance imaging (ULF-MRI) has attracted attention because of its low running costs and minimum patient exposure. An optically pumped atomic magnetometer (OPAM) is a magnetic sensor with high sensitivity in the low frequency range, which does not require a cryogenic cooling system. In an effort to develop a ULF-MRI, we attempted to measure the free induction decay MR signals with an OPAM. We successfully detected the MR signals by combining an OPAM and a flux transformer, demonstrating the feasibility of the proposed system.
机译:由于其运行成本低和最低患者暴露,超低场磁共振成像(ULF-MRI)引起了注意力。光学泵浦原子磁力计(OPAM)是低频范围内具有高灵敏度的磁传感器,其不需要低温冷却系统。为了开发ULF-MRI,我们试图测量使用OPAM的自由感应衰变MR信号。我们通过组合OPAM和磁通互感器成功检测MR信号,证明了所提出的系统的可行性。