ISSUES AND DESIGN SOLUTIONS ASSOCIATED WITH PERFORMING MRI SCANS ON PATIENTS WITH ACTIVE IMPLANTABLE MEDICAL DEVICES

机译:与活动植入医疗器械患者进行MRI扫描相关的问题和设计解决方案

获取原文

摘要

Magnetic Resonance Imaging (MRI) has become one of medicine's most important diagnostic tools. However, due to patient safety concerns, MRI is contraindicated by both device and MRI equipment manufacturers for patients with active implanted medical devices (AIMDs). The primary concern is overheating of implanted leadwires due to currents induced from the powerful RF fields of the MRI scanner. In pacemaker patients, heating of myocardial tissue has caused increase in pacemaker capture threshold and in some cases complete loss of capture (inability to pace). Permanent damage to an implantable cardioverter defibrillator and at least one patient death and another with severe burns along the wires of deep brain electrodes have also been reported. The authors, with pacemaker leadwires placed in a "worse case" MRI scan condition, have measured distal tip leadwire temperatures of up to 57 degrees C (more than sufficient to cause tissue damage). Another risk is localized myocardial ablation which could result in changes in the action potential vector during atrial/ventricular contraction. However, in contrast to reports of problems, there have been several recent anecdotal reports of MRI scans being safely performed on non-pacemaker dependant patients under highly controlled conditions. Proper diagnosis, treatment and management of a number of life threatening diseases such as cancer, neurological and brain disorders are made possible by MRI. Accordingly, the physician, with informed patient consent, must sometimes ignore the legal contraindications, weigh the risk factors, and go ahead and perform an MRI on an AIMD patient. This paper quantifies the attendant risks of performing MRI on AIMD patients and discusses means of mitigating certain hazards such as leadwire overheating including the performance a new leadwire distal tip resonant bandstop filter chip (MRI Chip) which presents a high impedance at the MRI pulsed RF frequency.
机译:磁共振成像(MRI)已成为药物最重要的诊断工具之一。然而,由于患者的安全问题,MRI由设备和MRI设备制造商进行禁忌,为有源植入医疗设备(AIMDS)患者。由于来自MRI扫描仪的强大RF场所引起的电流,主要关注的是植入的导线过热。在起搏器患者中,心肌组织的加热导致起搏器捕获阈值增加,并且在某些情况下,捕获的损失(无法速度)。还据报道对植入的心脏除颤器的永久性损害以及至少一个患者死亡,并且沿着深脑电极电线的严重燃烧。作者,带有起搏器的前线放置在“较差的情况下”MRI扫描条件下,具有高达57摄氏度的远端尖端引线温度(超过足以引起组织损伤)。另一种风险是局部心肌消融,其可能导致心房/心室收缩期间动作电位载体的变化。然而,与问题的报告相比,最近有几个关于在高度受控条件下对非起搏器依赖患者安全进行的MRI扫描的轶事报告。 MRI使得癌症,神经系统和脑疾病等许多危及危及疾病的适当诊断,治疗和管理。因此,具有明智的患者同意的医生必须忽视法律禁忌症,权衡风险因素,并继续前进并在AIMD患者上执行MRI。本文量化了对AIMD患者进行MRI进行MRI的伴随风险,并讨论了减轻了某些危害的方法,如引线过热,包括新的引线远端谐振带芯片滤波器芯片(MRI芯片),它在MRI脉冲RF频率下具有高阻抗。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号