【24h】

Development of a remote handheld cardiac arrhythmia monitor

机译:开发远程手持式心脏心律失常监测仪

获取原文

摘要

In this paper we present the design and development of a real-time remote handheld cardiac arrhythmic monitoring system (RCAM). A client-server model based on Internet protocols was used. ECG data was transmitted from the remote handheld client to a centralized server, where the QRS and premature ventricular contraction detection algorithms were implemented and graded depending on the number and pattern of PVCs present. The QRS sensitivity and specificity on ECG records from Physionet archives in absence of arrhythmia was 100% and 99.62%, while in presence of arrhythmia was 99.34% and 99.31%. The average 'negative time' measured on ventricular tachyarrhythmia records was 92 seconds. The RCAM can provide remote detection of cardiac abnormalities and give specific diagnosis and recommendations of actions to be taken immediately. The limitation due to the inability of the PDA to perform complex computations was overcome by the use of the remote server.
机译:在本文中,我们介绍了实时远程手持心性心律失常监测系统(RCAM)的设计和开发。使用基于Internet协议的客户端 - 服务器模型。将ECG数据从远程手持客户端发送到集中式服务器,其中实现QR和过早的心室收缩检测算法,并根据存在的PVC的数量和模式来实现和分级。在没有心律失常的物理仪档案中的QRS敏感度和心电图记录的特异性为100%和99.62%,而心律失常存在99.34%和99.31%。在心室性心律失常记录上测量的平均“负时间”为92秒。 RCAM可以提供远程检测心脏异常,并提供具体的诊断和立即采取行动的建议。通过使用远程服务器克服了PDA无法执行复杂计算导致的限制。