首页> 外文会议>Conference on Precision Electromagnetic Measurements >A novel hollow active auxiliary electrode applied for the vertical calculable cross-capacitor at NIM
【24h】

A novel hollow active auxiliary electrode applied for the vertical calculable cross-capacitor at NIM

机译:一种新型的中空有源辅助电极,用于NIM的垂直可计算交叉电容器

获取原文

摘要

In order to compensate the end-effect error in the vertical calculable cross-capacitor with one fixed and one movable guard electrodes, two different methods including the nosepiece and the modified active auxiliary electrode are usually adopted in the main worldwide national measurement institutes. However, the latter electrical approach will have some difficulties in assembling the mirrors for the resonant cavity of Fabry-Perot laser interferometer. This paper describes one novel hollow active auxiliary electrode which can satisfy the optical requirements and the model experimental results show that it has similar excellent compensation effect with the modified active auxiliary electrode.
机译:为了补偿具有一个固定电极和一个可移动保护电极的垂直可计算交叉电容器的最终效应误差,世界范围内的主要国家测量机构通常采用两种不同的方法,包括鼻架和改进的有源辅助电极。然而,后一种电学方法在组装用于法布里-珀罗激光干涉仪的谐振腔的反射镜时会遇到一些困难。本文介绍了一种新型的能满足光学要求的空心有源辅助电极,模型实验结果表明,该空心有源辅助电极具有与改性有源辅助电极相似的优良补偿效果。