The Effects of Radiometry on the Accuracy of Intensity Based Registration

机译:辐射测量术的影响对基于强度的注册精度

获取原文

摘要

Besides several other factors, radiometric differences between a reference and a floating image greatly influence the achievable accuracy of image registration. In this work we derive the magnitude of registration inaccuracy coming from changes in radiometric properties. This is done for the example of medical X-ray image registration. We therefore estimate the change of image intensity with respect to object shape, X-ray attenuation of the object material and the initial X-ray energy by modeling a simplified image formation process. The change in intensity is then used to determine a closed form estimation of the resulting registration error, independent from a specific registration algorithm. Finally the theoretical calculations are compared to the accuracy of intensity based registration performed on X-ray images with different radiometric properties. Results show that the herewith derived accuracy estimation is well suited to predict the achievable accuracy of a registration for images with radiometric differences.
机译:除了几个其他因素外,参考和浮动图像之间的辐射差异大大影响了图像配准的可实现的准确性。在这项工作中,我们推出了来自辐射性能变化的注册不准确性的大小。这是为医疗X射线图像配准的例子完成的。因此,我们通过建模简化的图像形成过程来估计对物体形状,对象材料的X射线衰减和初始X射线能量的图像强度的变化。然后使用强度的变化来确定所得注册误差的闭合形式估计,与特定的登记算法无关。最后,将理论计算与具有不同辐射特性的X射线图像执行的强度基于的注册的准确性进行比较。结果表明,这里的衍生精度估计非常适合预测具有辐射差异的图像的注册的可实现准确性。