Vicarious calibration of Terra/ASTER/VNIR with desert scenes together with cross calibration

机译:用沙漠场景对Terra / ASTER / VNIR进行替代性校准以及交叉校准

获取原文

摘要

Fourteen and half years of vicarious calibration and cross calibration for Terra/ASTER/VNIR (Visible and Near Infrared Radiometer, one of three ASTER mission instruments onboard Terra satellite) are summarized together with an error budget analysis. As the results, it is found that Onboard Calibration (OBC) data derived Radiometric Calibration Coefficient (RCC) is situated within a range of uncertainty of vicarious calibration except the period after approximately 4000 days after launch. In other word, uncertainty of OBC RCC is overlapped with uncertainty of vicarious calibration for about 4000 days since launch. Meanwhile, cross calibration with MODIS, MISR and ETM+ supports vicarious calibration of VNIR. In other word, Cross RCC shows much consistent trend to Vicarious RCC than OBC RCC.
机译:总结了Terra / ASTER / VNIR(可见和近红外辐射仪,Terra卫星上的三种ASTER任务仪器之一)经过十四年半的替代校准和交叉校准,并进行了误差预算分析。结果发现,除了发射后约4000天之后的时间以外,车载校准(OBC)数据衍生的辐射校准系数(RCC)处于替代校准的不确定性范围内。换句话说,自发射以来,OBC RCC的不确定性与替代校准的不确定性重叠了大约4000天。同时,使用MODIS,MISR和ETM +进行交叉校准可支持VNIR的替代校准。换句话说,与RBC RCC相比,Cross RCC向Vicarious RCC显示出一致的趋势。