首页> 外文会议>IEEE International Symposium on Biomedical Imaging >Automatic detection of head refixation errors in fractionated stereotactic radiotherapy (FSR)
【24h】

Automatic detection of head refixation errors in fractionated stereotactic radiotherapy (FSR)

机译:分馏立体定向放射治疗中的头原析误差的自动检测(FSR)

获取原文

摘要

Patient surface images are acquired using a novel 3D camera when the patient is at the CT-simulation position and after setup for fractionated stereotactic treatment. The simulation and treatment images are aligned through an initial registration using several feature points followed by a refined automatic matching process using an iterative-closest-point mapping-align algorithm. All of the video-surface images could be automatically transformed to the machine coordinate according to the calibration file obtained from a template image. Phantom tests have demonstrated that we can capture surface images of patients in a second with spatial resolution of submillimeter. A millimeter shift and one-degree rotation relative to the treatment machine can be accurately detected. The entire process takes about two minutes. Our primary result on patients involved in a clinical trial is very promising. This research is partially supported by INH Grant 1R43CA91690-01 and NIH CA88843.
机译:当患者处于CT仿真位置和分级立体定向处理后,使用新型3D相机获取患者表面图像。模拟和处理图像通过使用若干特征点对准,然后使用迭代 - 离点映射算法进行精细的自动匹配过程。根据从模板图像获得的校准文件,所有视频曲面图像都可以自动转换为机器坐标。 Phantom测试已经证明我们可以在亚瑟倍数钟的空间分辨率中捕获患者的表面图像。可以精确地检测毫米偏移和相对于处理机的一度旋转。整个过程大约需要大约两分钟。我们对参与临床试验的患者的主要结果非常有前途。通过INH授予1R43CA91690-01和NIHCA88843部分地支持该研究。