首页> 外文会议> >Automatic detection of head refixation errors in fractionated stereotactic radiotherapy (FSR)
【24h】

Automatic detection of head refixation errors in fractionated stereotactic radiotherapy (FSR)

机译:在立体定向放射治疗(FSR)中自动检测头部固定错误

获取原文
获取外文期刊封面目录资料

摘要

Patient surface images are acquired using a novel 3D camera when the patient is at the CT-simulation position and after setup for fractionated stereotactic treatment. The simulation and treatment images are aligned through an initial registration using several feature points followed by a refined automatic matching process using an iterative-closest-point mapping-align algorithm. All of the video-surface images could be automatically transformed to the machine coordinate according to the calibration file obtained from a template image. Phantom tests have demonstrated that we can capture surface images of patients in a second with spatial resolution of submillimeter. A millimeter shift and one-degree rotation relative to the treatment machine can be accurately detected. The entire process takes about two minutes. Our primary result on patients involved in a clinical trial is very promising. This research is partially supported by INH Grant 1R43CA91690-01 and NIH CA88843.
机译:当患者处于CT模拟位置时,以及在进行分级立体定向治疗后,将使用新型3D相机获取患者表面图像。通过使用多个特征点的初始配准,对模拟和处理图像进行对齐,然后使用迭代最近点映射对齐算法进行精确的自动匹配过程。根据从模板图像获得的校准文件,可以将所有视频表面图像自动转换为机器坐标。幻影测试表明,我们可以在一秒内以亚毫米的空间分辨率捕获患者的表面图像。可以准确地检测到相对于治疗机的毫米偏移和一度旋转。整个过程大约需要两分钟。我们对参与临床试验的患者的主要结果非常有希望。该研究得到了INH Grant 1R43CA91690-01和NIH CA88843的部分支持。