首页> 外文会议>International Conference on BioMedical Engineering and Informatics >Methods of interruption removal and pulse rate calculations in pulse wave signals based on wearable electronic eyeglasses
【24h】

Methods of interruption removal and pulse rate calculations in pulse wave signals based on wearable electronic eyeglasses

机译:基于可穿戴电子眼镜的脉搏波信号中的中断消除和脉率计算方法

获取原文
获取外文期刊封面目录资料

摘要

Baseline drift and other motion artifacts caused by breathing and physical movements need to be reduced from useful signals in pulse wave signals detection process, and the relevant effective methods to calculate pulse rate are also urgently needed. This paper is based on the wearable electronic eyeglasses aiming at measuring pulse signals with a three-axis acceleration sensor to catch body movement signals. It presents a removal method of the baseline drift using cubic spline interpolation, applies the least mean square (LMS) method which needs additional accelerometer signals in reducing the motion artifacts, and adopts the improved exceeding value method to detect the principal peaks and calculate the pulse rates of the pulse signals. Experiment results show that the combination of the three methods can effectively remove the baseline drift, reduce the motion artifacts, detect the principal peaks of pulse signals, and accurately calculate the pulse rates.
机译:在脉搏波信号检测过程中,需要从有用信号中减少由呼吸和身体运动引起的基线漂移和其他运动伪影,并且迫切需要用于计算脉搏率的相关有效方法。本文基于可穿戴电子眼镜,旨在通过三轴加速度传感器测量脉冲信号以捕获人体运动信号。提出了使用三次样条插值法消除基线漂移的方法,应用了最小均方(LMS)方法,该方法需要附加的加速度计信号来减少运动伪像,并采用改进的超值方法来检测主峰并计算脉冲脉冲信号的速率。实验结果表明,三种方法的结合可以有效消除基线漂移,减少运动伪像,检测脉搏信号的主峰,并准确计算脉搏率。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号