The paper presents a numerical worst-case study of the coupling between coagulators and cardiac pacemakers under consideration of the human body. For this purpose a special body model has been implemented which meets the geometrical dimensions given in DIN 33402. With this model whose electrical properties correspond to a weighted average of those of different tissues of the human body the influence of the latter can be taken into account in the computations. Different parameters such as position of the pacemaker and its electrode, coagulation frequency, and coagulation point on the surface of the human body are considered. Owing to the results of previous investigations simplified approaches to account for the dielectric coating of the coagulator electrodes and housing of the pacemaker can be used. The investigation shows that for the scenario described above peak-to-peak voltages up to about 20 mV result at the input port of the cardiac pacemaker.
机译:本文介绍了在考虑人体期间凝结器与心脏起搏器之间的偶联的数值最坏情况研究。 为此目的,已经实施了一种特殊的身体模型,其满足DIN 33402中给出的几何尺寸。通过这种模型,其电性质对应于人体不同组织的加权平均值可以考虑后者的影响 在计算中。 考虑不同的参数,例如起搏器的位置及其电极,凝血频率和人体表面上的凝固点。 由于先前研究的结果,可以使用简化的方法来解释用于起搏器的凝固器电极和壳体的介电涂层。 该研究表明,对于上述峰值到峰值电压的场景,在心脏起搏器的输入端口处高达约20 mV的结果。