首页> 外文会议>European Microwave Conference >Development of a 70 MHz unit for hyperthermia treatment of deep seated breast tumors
【24h】

Development of a 70 MHz unit for hyperthermia treatment of deep seated breast tumors

机译:一种70 MHz单位进行深热治疗深坐乳腺肿瘤的70 MHz

获取原文

摘要

Hyperthermia of tumors in intact breast extending beyond the heating depth of our superficial 434 MHz CFMA antennas requires an alternative approach. A dedicated system was designed for this purpose, consisting of a treatment bed fitted with a 50×40×16 cm temperature controlled open water bolus. The patient lies in prone position with the breast immersed in the water positioned in front of a 34×20 cm 70 MHz waveguide placed in the bottom of the bolus. Hyperthermia was applied once a week for the whole breast with the 70 MHz applicator for 6 patients treated with thermoradiotherapy for deep lesions of recurrent breast cancer or melanoma. Two 14-sensor thermocouple thermometry probes were placed in catheters to monitor the invasive temperature. Results: The combination of 300-900 W antenna power and a water temperature of 42°C was well tolerated for the entire session of one hour and resulted in good tumor temperatures with T10 = 42.2°C, T50 = 41.1°C, T90 = 39.8°C. No toxicity or complaints were associated with the heating. A water mattress and other measures were needed to assure a comfortable position throughout the treatment. Conclusion: the 70 MHz breast applicator system performed well and tumor temperatures were good.
机译:完整乳房的肿瘤的热疗超出了我们肤浅的434 MHz CFMA天线的加热深度需要替代方法。专用系统为此目的而设计,由配有50×40×16cm温控开放式水推注的处理床组成。患者在静止的乳房浸入位于34×20cm 70MHz波导的水中放置在推注的34×20cm 70MHz波导的水中的倾倒位置。全乳液每周施用一次热疗,用70 MHz涂抹器,用于6名患者治疗热轧疗法的经常性乳腺癌或黑色素瘤。将两个14传感器热电偶测压探针置于导管中以监测侵入性温度。结果:300-900 W天线功率的组合和42℃的水温在整个一半的整个会议上耐受良好耐受,导致T10 = 42.2°C,T50 = 41.1°C,T90 = 39.8°C。没有毒性或抱怨与加热有关。需要水床垫和其他措施来确保整个治疗过程中的舒适位置。结论:70 MHz乳房涂抹器系统表现良好,肿瘤温度都很好。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号