A New Algorithm of Drop Speed Measuring of Infusion Monitoring Device

机译:输液监控装置滴速测量的新算法

获取原文

摘要

Alarm accuracy of infusion monitoring device mostly depends on drop speed measuring of infusion. A new drop speed calculation is put forward for the discontinuous characteristics of infusion. 5 drops is taken as a calculation unit. When new drop is detected, the time interval between new drop & its previous one will replace that between first one drop and second drop. The average drop speed will be calculated like that. Compared with the drop speed mentioned in Reference, this advantage of algorithms is of real time, reliability and effectively reducing wrong-alarm rate.
机译:输液监控装置的报警精度主要取决于输液滴速的测量。针对输液的不连续性,提出了一种新的滴速计算方法。以5滴为计算单位。当检测到新的墨滴时,新墨滴与其之前的墨滴之间的时间间隔将替换第一墨滴和第二墨滴之间的时间间隔。平均下降速度将像这样计算。与参考中提到的下降速度相比,该算法的优点是实时,可靠并有效降低了错误警报率。