We present the combination of high speed video monitoring, force measurement and Optical Coherence Tomography (OCT) as a new method for the characterization of liquid jets for needle-free tissue injection applications. It is well known that physical parameters, such as jet velocity and jet diameter determine the penetration characteristics of liquid micro-jets into a specific tissue [1]. The optimum injection parameters vary significantly for different types of tissues. Therefore, we have used high speed video monitoring and force sensor measurements to characterize the liquid jet. The penetration into the tissue is then analyzed by Optical Coherence Tomography (OCT) [2]. This enables the general description of the specific jet's physical properties as well as the penetration characteristics with respect to a specific tissue. In the experimental study, jets with diameters of 100 μm and jet lengths from 6 to 18 ms (∼ 2 to 9 μl) have been injected into colon tissue from pigs.
机译:我们介绍高速视频监控,力测量和光学相干断层扫描(OCT)的组合,作为表征无针组织注射应用中液体射流的一种新方法。众所周知,物理参数(例如射流速度和射流直径)决定了液体微射流进入特定组织的渗透特性[1]。对于不同类型的组织,最佳注射参数会有很大差异。因此,我们使用了高速视频监控和力传感器测量来表征液体射流。然后通过光学相干断层扫描(OCT)分析进入组织的穿透力[2]。这使得能够总体描述特定射流的物理特性以及相对于特定组织的穿透特性。在实验研究中,直径为100μm的喷嘴和长度为6到18 ms(约2到9μl)的喷嘴已从猪的结肠组织中注入。