Mathematical derivation of calculating blood flow volume (BFV) at arteriovenous fistula (AVF) using a newly-developed photoplethysmography (PPG) sensor is presented in this work. Also, the readout circuit of the PPG sensor intended to increase the signal-noise ratio (S/NR) is designed and presented in this work. The designed PPG sensor equipped with derived mathematical equations shows high correlation (R2 = 0.7563) and low error (RMSE = 212 ml/min) as compared to the gold standard.
机译:在这项工作中,提出了使用新型显影性的光增生术(PPG)传感器在动血管瘘(AVF)处计算血流量(BFV)的数学衍生。此外,在该工作中设计和呈现了旨在增加信噪比(S / NR)的PPG传感器的读出电路。配备有衍生数学方程的设计的PPG传感器显示出高相关(R 2 = 0.7563),与金标准相比,低误差(RMSE = 212毫升/分钟)。