首页> 外文会议>IASTED international conference on Biomedical Engineering >A LOW-COST MICROCONTROLLER BASED GAS FLOWMETER USING A SPECIALLY DESIGNED PARABOLIC AIRWAY RESISTOR
【24h】

A LOW-COST MICROCONTROLLER BASED GAS FLOWMETER USING A SPECIALLY DESIGNED PARABOLIC AIRWAY RESISTOR

机译:使用专门设计的抛物线气道电阻器的低成本微控制器的气体流量计

获取原文

摘要

A parabolic airway resistor is designed and used as a flowmeter in the periphery of a microcontroller, to measure air and O2 flows, in the range 0-50 lpm, in medical device applications . The accuracy of the readings is 2% at maximum and the reproducibility is better than 0.05% in the range 5-50 lpm.. The accuracy for flow measurements below 5 lpm. can be improved by using a different airway resistor with a smaller orifice diameter. The cost of the flowmeter is around 250 $.
机译:抛物线气道电阻设计并用作微控制器周边的流量计,以测量空气和O2在医疗设备应用中的0-50毫升范围内的流量。最大读数的准确性为2%,再现性优于0.05%,范围为5-50液化因素..流量测量的精度低于5升。可以通过使用具有较小孔口直径的不同气道电阻来改进。流量计的成本约为250美元。