Full-duplex communication is realized by using ASK modulation for downlink (the channel from controller to the implant) and by using FSK modulation for uplink (from implant to the controller). The ASK modulation is done at the remote powering frequency and by using the inductive link. The classE type power amplifier is fed by the modulated data. For the uplink communication a stand alone transmitter is designed with a crystal oscillator. The implant node is modeled by using a microcontroller which is sending data and a dummy resistor to mimic the power consumption of the sensors. This microcontroller is also powered up by remote powering.
机译:通过使用向下行链路(来自控制器到植入器的通道)的询问调制来实现全双工通信,并通过使用FSK调制来实现上行链路(从植入物到控制器)。询问调制以远程供电频率进行,使用电感链路完成。 CLASSE型功率放大器由调制数据馈送。对于上行链路通信,独立的发送器设计有晶体振荡器。通过使用正在发送数据和虚设电阻以模拟传感器的功耗的微控制器来建模植入节点。此微控制器也通过远程供电。