The aim of this paper is to characterize the response of radiochromic film GAFCHROMIC - RV2 with the radiation dose using a transmission densitometer (McBeth), reflection densitometer (X-Rite) and flatbed scanner HP Scanjet 3500c. The sensitometric curve relating irradiated film's Optical Density to the patient absorbed Dose is obtained. This curve can be later employed to have an estimate of the absorbed dose by the patient undergoing a radiological procedure. Finally, the precision of the curve is investigated by simulating real cases. Results of this test show how obtaining the Optical Density values by means of scanner, although a more complex procedure, gives more precise results (less than 4%) than by means of densitometers employed (up to 20%).
机译:本文的目的是使用透射密度计(McBeth),反射密度计(X-Rite)和平板扫描仪HP Scanjet 3500c来表征辐射致变色膜GAFCHROMIC-RV2的辐射剂量响应。获得将辐照膜的光密度与患者吸收剂量相关的灵敏度曲线。以后可以使用该曲线来估计正在接受放射治疗的患者吸收的剂量。最后,通过模拟实际情况来研究曲线的精度。该测试的结果表明,尽管扫描程序较复杂,但如何通过扫描仪获得光密度值,却比使用密度计(高达20%)获得更精确的结果(小于4%)。