首页> 外文会议>Latin American congress on biomedical engineering >Criterios Cuantitativos de Selección de la Bobina Receptora de Radiofrecuencia para Estudiar Animales Pequeños en Equipos Clínicos de Resonancia Magnética
【24h】

Criterios Cuantitativos de Selección de la Bobina Receptora de Radiofrecuencia para Estudiar Animales Pequeños en Equipos Clínicos de Resonancia Magnética

机译:在临床磁共振设备中​​选择用于研究小型动物的射频接收线圈的量化标准

获取原文

摘要

The purpose of the study was to select the best radiofrequency coils for the study of small animal models in 1,5T MRI whole body clinical equipment. The signal -to-noise ratio (SNR), the contrast-to- noise ratio (CNR) and the nom uniformities (NUI) are evaluated of phantom using two different radiofrequency coils: the Head and Flex S from Symphony machine (Siemens). It is demonstrated that the SNR, CNR and NUI are better with the flexible coil. As it is expected, that in the two phantom experiments the SNR for each phantom is increasing more than 25 %. The NUI decrease from the coils with one phantom to the same coil with two phantom. This results allow greater productivities and reproducibility of animals models studies in MRI clinical machine. As example, the T2 and DWI images obtained in this work of Winstar rats are showed.
机译:该研究的目的是为研究1,5T MRI全身临床设备中的小动物模型选择最佳的射频线圈。使用两个不同的射频线圈:Symphony机器(Siemens)的Head和Flex S对幻像进行了信噪比(SNR),对比度对噪声比(CNR)和标称均匀性(NUI)评估。结果表明,使用柔性线圈时,SNR,CNR和NUI更好。不出所料,在两个体模实验中,每个体模的SNR都增加了25%以上。 NUI从具有一个幻像的线圈减少到具有两个幻像的相同线圈。该结果允许在MRI临床机器中进行动物模型研究的更高的生产率和可重复性。例如,显示了在Winstar大鼠的这项工作中获得的T2和DWI图像。