首页> 外文会议>International Conference on Advancements in Nuclear Instrumentation Measurement Methods and their Applications >NEMA NU4#x2013;2008 performance evaluation for the microPET FOCUS 120 and iodine-124
【24h】

NEMA NU4#x2013;2008 performance evaluation for the microPET FOCUS 120 and iodine-124

机译:NEMA NU4– 2008年微量心焦点120和碘-124的性能评估

获取原文

摘要

This study aimed to evaluate the performance of the microPET FOCUS 120 in terms of counting rate capability and image quality when using 124I. The noise equivalent count rate (NECR) index was defined as: NECR = RT2/RP + RGP (T = true, P = prompt, GP = gamma prompt). The maximum NECR (48 kcps) value was obtained for the 250-590 keV energy window (EW) with 6 ns time window (TW). The two other EW (350-590 and 350-650 keV) gave comparable maximum NECR of 43 kcps for the 6 ns TW. The 2 ns TW reduced the NECR by 45 to 50 % Scatter and gamma plus scatter fractions were almost not TW dependent. The 250-590 keV window showed the largest scatter fraction with 23% For the NEMA NU4-2008 image quality phantom, coefficients of variation in the uniform part increased from about 8 to 12% when decreasing TW from 6 to 2 ns. The lowest spill-over-ratios (SOR) for the non emitting regions were obtained for the 350-590 and 350-650 keV EW. Recovery coefficients (RC) of the hot rods were the highest for the 350-590 keV EW and 2 ns TW. Scatter correction led to a decrease in RC and occasionally (with FBP or 3DRP) to a negative value for the SOR. The combination of the 350-590 keV energy window with 6 ns time window appeared to be a good compromise between counting rate capability and image quality for the FOCUS 120, especially when MAP reconstruction was used. The scatter correction algorithm provided by the constructor should be used with caution.
机译:本研究旨在评估Micropet焦点120在使用124i时计算速率能力和图像质量方面的性能。噪声等效计数率(NECR)索引被定义为:neCR = RT2 / RP + RGP(T = TRUE,P =提示,GP = Gamma提示)。为250-590 kev能量窗口(EW)获得了最大NECH(48 KCPS)值,具有6个NS时间窗口(TW)。另外两个EW(350-590和350-650 kev)为6 NS TW提供了43 KCP的相当最大NECR。将2NS T降低45至50%的散射,γ加散射级分几乎没有依赖于TW。 250-590 keV窗口显示NEM4-2008图像质量模拟的23%的最大散射分数,均匀部分的变化系数从6到2 ns减小时增加了约8至12%。为350-590和350-650 keV EW获得了非发光区域的最低溢出比率(SOR)。热棒的恢复系数(RC)对于350-590 kev Ew和2 ns tw最高。散点校正导致RC的减少,偶尔(用FBP或3DRP)对SOR的负值。 350-590 Kev能量窗口的组合具有6个ns时间窗口的计数率能力与焦点120的计数率能力和图像质量之间的良好折衷,尤其是在使用地图重建时。应谨慎使用构造函数提供的分散校正算法。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号