The aim of this study was to measure the concentration of beta2-microglobulin (B2M) in the spent dialysate using optical method utilizing fluorescence spectra of the spent dialysate. Eight uremic patients from Linkoping were studied during their three dialysis treatments in one week at the Department of Dialysis and Nephrology at Linkoping University Hospital. Dialysate samples were taken during each treatment and analyzed, B2M concentration was measured in laboratory, and fluorescence spectra was measured with spectrofluoropho-tometer. The fluorescence spectral values were transformed into B2MG concentration using regression model from total material noted as fluorescence method (F). Achieved results were compared regarding mean values and SD.Mean value of B2M measured in laboratory was 1.71±1.0 mg/1 and by F 1.72±0.9 mg/1. Concentrations were not significantly different (p≤0,05).This study indicates that it is possible to estimate the concentration of B2M using only fluorescence values of the spent dialysate.
机译:这项研究的目的是使用光学方法利用废透析液的荧光光谱来测量废透析液中β2-微球蛋白(B2M)的浓度。在林雪平大学医院透析和肾脏病科,一周内对八名来自林雪平的尿毒症患者进行了三次透析治疗。在每次处理期间采集透析液样品并进行分析,在实验室中测量B2M浓度,并使用分光光度计测量荧光光谱。荧光光谱值使用总模型的回归模型转换为B2MG浓度,记为荧光法(F)。比较平均值和SD的结果。实验室测得的B2M平均值为1.71±1.0 mg / 1,F为1.72±0.9 mg / 1。浓度无显着差异(p≤0.05)。这项研究表明,仅使用废透析液的荧光值即可估算B2M的浓度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号