The presence of the rib cage is a significant hindrance to the clinical applicability of focused ultrasound surgery for various internal organs, including liver, kidney, and pancreas. The main problems associated with the rib cage are considered to be: the strong reflections that occur at the soft tissue/rib interfaces; the highly attenuating effect of the ribs; and the scattering and diffraction that occur at the ribs boundaries. In this study, a 2D high speed ultrasound scanning system was used to directly measure the acoustic field in the focal plane after propagation of focused ultrasound through an acoustic absorber rib-mimicking material and an epoxy rib-mimicking material. Numerical simulations were also performed with finite element analysis software (PZFlex, Weidlinger Associates, USA) to investigate the changes in the acoustic field with different anatomical configurations. Briefly, due to the defocusing nature of the rib-mimics, the acoustic pressure level at the focus was substantially reduced. This study quantitatively analyzed the effect of focus splitting associated with the ribs to provide reference information for further clinical applications.
机译:肋骨笼的存在严重阻碍了聚焦超声手术在包括肝脏,肾脏和胰腺在内的各种内部器官的临床应用。与肋骨保持架相关的主要问题被认为是:在软组织/肋骨界面处发生的强烈反射;肋骨的高度衰减作用;以及在肋骨边界处发生的散射和衍射。在这项研究中,使用二维高速超声扫描系统直接测量聚焦超声通过吸声肋模仿材料和环氧肋模仿材料传播后的焦平面中的声场。还使用有限元分析软件(PZFlex,Weidlinger Associates,美国)进行了数值模拟,以研究不同解剖结构的声场的变化。简而言之,由于模拟肋的散焦特性,焦点处的声压级大大降低了。这项研究定量分析了与肋骨相关的焦点分裂的影响,为进一步的临床应用提供参考信息。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号