Measurement of prostate viscoelasticity using Shearwave Dispersion Ultrasound Vibrometry (SDUV): An in vitro study

机译:使用Shearwave分散超声振动法(SDUV)测量前列腺粘弹性:一项体外研究

获取原文

摘要

We report shear stiffness and viscosity “virtual biopsy” measurements of three excised non-cancerous human prostates using a new characterization tool known as Shearwave Dispersion Ultrasound Vibrometry (SDUV) in vitro. Improved methods for prostate guided-biopsy are required to effectively guide needle biopsy to the suspected site. In addition, tissue stiffness measurement helps identifying a suspected site to perform biopsy because stiffness has been shown to correlate with pathologies such as cancerous tissue. More importantly, early detection of prostate cancer may guide minimally-invasive therapy and eliminate insidious procedures. In this work, “virtual biopsies” were taken in multiple locations in three excised prostates; SDUV shear elasticity and viscosity measurements were performed at the selected “suspicious” locations within the prostates. We found that the range of the shear elasticity and viscosity were 1.31–12.81 kPa and 1.10–6.82 Pa⋅s, respectively. The mean and standard deviations for the shear elasticity and viscosity were 5.20 ± 2.55 kPa and 2.68 ± 1.47 Pa⋅s, respectively. SDUV measurements of prostate elasticity and viscosity are generally in agreement with preliminary values reported previously in the literature. It is however important to emphasize here that the obtained viscoelastic parameters values are local, and not a mean value for the whole prostate.
机译:我们报告了三个新切除的非癌性人类前列腺的剪切刚度和粘度“虚拟活检”测量结果,该新前列腺癌的表征工具是使用新的表征工具,即体外剪切波分散超声振动测定法(SDUV)。需要用于前列腺穿刺活检的改进方法,以有效地将穿刺活检引导至可疑部位。此外,组织刚度测量有助于确定可疑部位以进行活检,因为已显示刚度与诸如癌性组织等病理相关。更重要的是,前列腺癌的早期发现可以指导微创治疗并消除阴险手术。在这项工作中,在三个切除的前列腺的多个位置进行了“虚拟活检”。 SDUV剪切弹性和粘度测量是在前列腺内选定的“可疑”位置进行的。我们发现剪切弹性和粘度范围分别为1.31–12.81 kPa和1.10–6.82 Pa·s。剪切弹性和粘度的平均偏差和标准偏差分别为5.20±2.55 kPa和2.68±1.47 Pa·s。 SDUV对前列腺弹性和粘度的测量通常与先前文献中报道的初步值一致。然而,在此必须强调的是,所获得的粘弹性参数值是局部的,而不是整个前列腺的平均值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号