Direct, In Vivo Calibration Measurement of ~(241)Am in the Axillary Lymph Nodes Using the Livermore Thoracic Phantom

机译:使用利弗莫尔胸腔幻影对腋窝淋巴结中〜(241)Am进行直接,体内校准测量

获取原文

摘要

The Livermore Torso Phantom, the consensus calibration phantom for measuring internally depositedradioactive materials in the lungs, liver and tracheo-bronchial lymph nodes, has been modified for calibratingdirect, in vivo measurements of ~(241)Am in the axillary lymph nodes. The axillary lymph nodes may containradioactive material mobilized from a wound in a finger or hand which may confound results of a direct, in vivomeasurement of the lungs or liver. Modifications to the Livermore Torso Phantom involved milling a pair ofparallel, flat bottom cylindrical holes into the left and right shoulders (below the humeral head) for inserting apair of solid 1.40 cm diameter cylindrical rods made of a 100% muscle tissue substitute. A precisely knownquantity of radioactive material was sealed in a 1 cm diameter, 2.54 cm deep well located at the end of each rod.The location adopted for the simulated lymph nodes was determined according to the position of the Level I andthe combined Level II +III axillary lymph nodes reported in the literature. Discrete calibration measurementswere performed for axillary lymph nodes located in the right and left side of the thorax using a pair of 3000 mm~2high resolution germanium detectors . The efficiency for measuring ~(241)Am in the right and left lymph nodes is2.603 % ± 0.0342 % and 5.453 % ± 0.0716 %, respectively. Activity deposited in the right and left axillarylymph nodes was found to contribute 0.183% and 0.289 %, respectively, to a lung measurement and 0.013 % and0.002 %, respectively, to a liver measurement. Thus, results of this research demonstrate that ~(241)Am deposited inthe axillary lymph nodes has the potential to confound direct, in vivo measurements of activity in the lung orliver.
机译:Livermore Torso幻影,用于测量内部沉积物的通用校准体模 修改了肺,肝和气管支气管淋巴结中的放射性物质,以进行校准 直接,体内测量腋窝淋巴结中的〜(241)Am。腋窝淋巴结可能含有 从手指或手的伤口中调出的放射性物质,可能会混淆直接体内的结果 肺或肝的测量。对Livermore Torso Phantom的修改涉及铣削一对 在左右肩部(肱骨头下方)平行,平坦的底部圆柱孔,用于插入 直径为1.40厘米的实心圆柱棒,由100%肌肉组织替代品制成。一个精确的已知 将一定量的放射性物质密封在位于每个棒末端的直径1厘米,深2.54厘米的井中。 模拟淋巴结的位置根据I级和 文献报道II级+ III级合并腋窝淋巴结肿大。离散校准测量 用一对3000 mm〜2的距离对位于胸腔左右两侧的腋窝淋巴结进行检查 高分辨率锗探测器。在左右淋巴结中测量〜(241)Am的效率为 分别为2.603%±0.0342%和5.453%±0.0716%。活动沉积在右腋窝和左腋窝 发现淋巴结分别占肺脏测量值的0.183%和0.289%,而占肺脏测量值的0.013%和0.289%。 分别为肝脏测量值的0.002%。因此,这项研究的结果表明〜(241)Am沉积在 腋窝淋巴结有可能混淆体内直接测量的肺部或肺部活动 肝脏。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号