首页> 外文会议>International conference on experimental mechanics >Image Quality Comparison of Computed Radiography and Digitized Film Radiograph
【24h】

Image Quality Comparison of Computed Radiography and Digitized Film Radiograph

机译:X光片与数字化胶片的影像质量比较

获取原文

摘要

The comparative study on the image quality between Computed Radiography (CR) and digitized film radiograph has been carried out. We have successfully digitized a radiographic film of an aluminum step wedge, an ASTM and coins using a flatbed scanner. Then, a comparative study was developed using CR as a reference. We used 4 set-up variations of kV and mA. The image quality was assessed based on the visual analysis: brightness, sharpness, and contrast, and the physical information analysis. The results show that visually the image quality of CR was better to the digitized film radiographs. However, CR was unable to distinguish the physical information such as the effect of voltage or current and the object characteristic to the image quality.
机译:已经进行了计算机射线照相(CR)和数字化胶片照相的图像质量的比较研究。我们已经使用平板扫描仪成功地数字化了铝台阶楔形物,ASTM和硬币的射线照相胶片。然后,以CR为参考进行了比较研究。我们使用了kV和mA的4种设置变化。根据视觉分析评估图像质量:亮度,清晰度和对比度,以及物理信息分析。结果表明,在视觉上CR的图像质量优于数字化胶片X射线照片。但是,CR无法区分物理信息,例如电压或电流的影响以及物体特性对图像质量的影响。