首页> 外文会议>32nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Safety evaluation of magnetic catheter steering with upgraded magnetic resonance imaging system
【24h】

Safety evaluation of magnetic catheter steering with upgraded magnetic resonance imaging system

机译:升级版磁共振成像系统对电磁导管转向的安全性评估

获取原文

摘要

Catheter navigation and placement through the arterial network is a major limitation for clinical procedure. In this article, a specific catheter tip and a modified clinical MRI scanner with an upgraded gradient system are used to steer a catheter through a single Y-shaped bifurcation. Safety aspects are analyzed to avoid the peripheral nerve stimulation (PNS) according to an empirical law of magnetostimulation and the magnetic field of upgraded 3D gradient coils. For a rabbit-sized device, the rising time of gradients system have to be limited to 1.7ms at 400mT.m−1 to avoid PNS. These rise time values allow the use of this system for catheter steering and other more demanding applications.
机译:通过动脉网络进行导管导航和放置是临床程序的主要限制。在本文中,使用了特定的导管尖端和带有升级的梯度系统的改良型临床MRI扫描仪来引导导管通过单个Y形分叉。根据磁刺激的经验定律和升级的3D梯度线圈的磁场,对安全方面进行了分析,以避免周围神经刺激(PNS)。对于兔子大小的设备,梯度系统的上升时间必须在400mT.m -1 上限制为1.7ms,以避免PNS。这些上升时间值允许将该系统用于导管操纵和其他更苛刻的应用。