A phoropter is one of the most popular ophthalmic instruments used in current optometry practice. The quality and verification of the instrument are of the utmost importance. In 1997, International Organization for Standardization published the first ISO standard for requirements of phoropters. However, in China, few standard and test method are suggested for phoropters. Research work on test method for phoropters was carried out early in 2004 by China National Institute of Metrology. In this paper, first, structure of phoropters is described. Then, theoretical considerations for its optical design are analyzed. Next, a newly developed instrument is introduced and measurements are taken. By calibration, the indication error of the instrument is not over 0.05m~(-1). Finally, measurement results show that the quality situation of phoropters is not as good as expected because of production and assembly error. Optical design shall be improved especially for combinations of both spherical and cylindrical lenses with higher power. Besides, optical requirements specified in ISO standard are found to be a little strict and hard to meet. A proposal for revision of this international standard is drafted and discussed on ISO meeting of 2007 held in Tokyo.
机译:验光仪是当前验光实践中使用的最流行的眼科仪器之一。仪器的质量和验证至关重要。在1997年,国际标准化组织发布了第一个针对phoropters要求的ISO标准。但是,在中国,对屈光力仪的标准和测试方法很少提出建议。中国国家计量学会于2004年初开展了验光仪测试方法的研究工作。在本文中,首先描述了phoropters的结构。然后,分析了其光学设计的理论考虑。接下来,介绍了一种新开发的仪器并进行了测量。经校准,仪器的指示误差不超过0.05m〜(-1)。最后,测量结果表明,由于生产和组装错误,phoropters的质量状况不如预期的好。尤其对于球面和圆柱透镜的组合,应改进光学设计,以提高功率。另外,发现ISO标准中规定的光学要求有些严格并且难以满足。在东京举行的2007年ISO会议上起草并讨论了修订此国际标准的提案。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号