机译:钛板在口腔颌面外科手术中的失败

获取原文

摘要

The failure of titanium bane plates for oral and maxillo-facial surgery during use was investigated. Microstructural examination of a titanium plate revealed the presence of equiaxed α grains and intergranular platelets, which were identified as β phase. Fractographic examination revealed that fracture happened by a transgranular cleavage mechanism associated with secondary intergranular cracking (brittle fracture). Selective attack was observed to occur on the surfaces of the implant. These results indicated that the premature fracture of the miniplate was caused by hydrogen embrittlement.
机译:研究了钛合金板在使用过程中在口腔和颌面部外科手术中的失败。钛板的微结构检查显示存在等轴的α晶粒和晶间血小板,它们被鉴定为β相。断口检查显示,骨折是由与继发性晶间裂纹(脆性断裂)相关的经晶裂机制引起的。观察到选择性攻击发生在植入物的表面上。这些结果表明,小板的过早断裂是由氢脆引起的。