An infrared tunable diode laser absorption spectrometer measuring trace moisture in air has been compared against the Low Frost-point Generator (LFG) of the UK National Physical Laboratory (NPL). The comparison between the two instruments followed on from earlier work described elsewhere at this conference, and took place during the course of the validation of the LFG in a trial phase of its operation down to frost points as low as -100°C (volume fraction of about 15 parts per billion (ppb)). In the form tested, the spectrometer performed with a resolution of better than 2 ppb, and demonstrated high linearity relative to the LFG, making it a highly useful tool for validating the performance of the generator. Conversely, the LFG measurements gave valuable information on the accuracy and repeatability of the spectrometer. The spectrometer and the LFG showed combined reproducibility of better than 12 ppb over a time period of three months. The measurements also demonstrated that the spectrometer had a significant offset of 215 ppb. A possible cause is thought to be captive water vapour within the photodetector of the spectrometer.
机译:红外可调谐二极管激光吸收光谱仪可测量空气中的微量水分,并已与英国国家物理实验室(NPL)的低霜点发生器(LFG)进行了比较。两种仪器之间的比较是根据本次会议其他地方所述的早期工作进行的,并且是在LFG的验证过程中进行的,试验的阶段是低至-100°C的霜点(体积分数)。约有十亿分之十五(ppb)的体积)。在测试的形式中,光谱仪的分辨率优于2 ppb,并且相对于LFG具有较高的线性度,这使其成为验证发生器性能的非常有用的工具。相反,LFG测量提供了有关光谱仪准确性和可重复性的有价值的信息。光谱仪和LFG在三个月的时间内显示出优于12 ppb的组合重现性。测量结果还表明,该光谱仪的显着偏移为215 ppb。可能的原因被认为是光谱仪的光电检测器内的水汽被俘获。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号