【24h】

Single Crystal Spectrometer FOX at KENS

机译:KENS的单晶光谱仪FOX

获取原文

摘要

Single crystal spectrometer FOX installed at H1 thermal netron line of KENS has been renewed recently for the measurement of very weak scattering. We have installed a multidetector system of 36 linearly placed ~3He detectors with collimators instead of former four-circle diffractometer and scintillator detectors. Though the system is quite simple, a large two-dimensional reciprocal space is observed effectively with high S/N rate on new FOX.
机译:最近更新了安装在KENS H1热内衬线的单晶光谱仪FOX,用于测量极微弱的散射。我们已经安装了由36个线性放置的〜3He探测器和准直仪组成的多探测器系统,而不是以前的四圆衍射仪和闪烁体探测器。尽管该系统非常简单,但是在新型FOX上可以以较高的信噪比有效观察到较大的二维倒数空间。