We will describe recent improvements to the SPEAR reflectometer at the Manuel Lujan Jr. Neutron Scattering Center at Los Alamos. One of the changes consists of wider convergent, incident-beam, collimation to take advantage of optical imaging for specular scattering. In addition, the instrument now views a partially coupled liquid hydrogen moderator as opposed to the decoupled moderator that was previous in-place. While the wavelength distribution is poorer, it matches the time (wavelength) resolution of the reflectometer more closely with the angular resolution.