During the process development of a polyester alloy medical device, a major relationship between product performance (incidence of failures) and the Theological properties was discovered. The Theological parameters were in turn found to be dependent on several factors during the polymer manufacturing, alloy compounding, and injection molding processes. It was also found, that controls far beyond the industry accepted standards were required for optimum quality. Through the implementation of a modified ASTM melt flow procedure at multiple points of the manufacturing process, consistent high quality production was achieved.
机译:在聚酯合金医疗器械的开发过程中,发现了产品性能(故障发生率)与神学特性之间的主要关系。反过来,发现神学参数取决于聚合物制造,合金复合和注塑过程中的几个因素。还发现,要获得最佳质量,就需要远远超出行业公认标准的控制。通过在制造过程的多个点实施改进的ASTM熔体流动程序,可以实现始终如一的高质量生产。