首页> 外文会议>World congress on medical physics and biomedical engineering;International congress of the IUPESM >Ankle and Knee Joint Angle Measurements during Gait with Wearable Sensor System for Rehabilitation
【24h】

Ankle and Knee Joint Angle Measurements during Gait with Wearable Sensor System for Rehabilitation

机译:使用可穿戴康复系统的步态过程中脚踝和膝关节角度的测量

获取原文

摘要

A method of joint angle measurement during gait using wearable sensors for rehabilitation was studied in this paper. The method corrected joint angles measured by gyroscopes using joint angles measured by accelerometers with Kalman filter. Although gyroscopes could measure joint angles, their offset drift caused error in measurement of joint angles. At first, we made up small sensor units consist of a gyroscope and an accelerometer. Then, the method was validated with the developed sensor units in measurement of ankle and knee joint angles of three healthy subjects under walking on short distance pathway and on treadmill for long time. The measured joint angles were compared with reference joint angles measured with optical motion measurement system simultaneously. The result for short distance walking showed low RMS errors and high correlation coefficients (3.19deg and 0.918 for the ankle joint, 2.98deg and 0.993 for the knee joint in average). The result for treadmill walking also showed low RMS errors and high correlation coefficients (3.04deg and 0.960 for the ankle, 4.19deg and 0.994 for the knee in average). Although the measurement accuracy decreased in some of trials of a specific subject because of sensor attachment position, the experimental results suggested that joint angle could be measured with good accuracy independent of measurement period of time, walking speeds and subjects.
机译:本文研究了一种使用可穿戴式康复传感器在步态中测量关节角度的方法。该方法使用带卡尔曼滤波器的加速度计测量的关节角度校正了陀螺仪测量的关节角度。尽管陀螺仪可以测量关节角度,但它们的偏移漂移会导致关节角度的测量误差。首先,我们组成了由陀螺仪和加速度计组成的小型传感器单元。然后,该方法在开发的传感器单元在短距离路径和跑步机上长时间步行下测量三个健康受试者的踝关节和膝关节角度时得到了验证。将测得的关节角度与同时用光学运动测量系统测得的参考关节角度进行比较。短距离步行的结果显示出较低的RMS误差和较高的相关系数(踝关节平均为3.19deg和0.918,膝关节平均为2.98deg和0.993)。跑步机行走的结果还显示出较低的RMS误差和较高的相关系数(踝部平均为3.04deg和0.960,膝部平均为4.19deg和0.994)。尽管在某些特定受试者的试验中,由于传感器的安装位置而导致测量精度降低,但实验结果表明,可以独立于测量时间,步行速度和受试者而以良好的精度测量关节角度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号