首页> 外文会议>World congress on medical physics and biomedical engineering;International congress of the IUPESM >Signal waveform agreement between spirometer and impedance pneumography of six chest band electrode configurations
【24h】

Signal waveform agreement between spirometer and impedance pneumography of six chest band electrode configurations

机译:肺活量计和六种胸带电极配置的阻抗肺图检查之间的信号波形一致性

获取原文

摘要

There is currently no approved clinical method for monitoring of pulmonary flow and volume without highly obtrusive instrumentation like a mask. Impedance pneumography (IP) seems to be the most promising method to establish accurate Holter type breathing measurement We measured IP and spirometer volume signal from 14 subjects in static conditions from six different tetrapolar electrode configurations using a chest band with textile silvertlectrodes. We compared the continuous waveforms from spirometer and IP with Pearson linear correlation coefficients p and found clear differences between the six configurations. For the best configuration the difference between the impedance and spirometer signals at p > 0.98 was almost insignificant for all subjects. In these static conditions the IP method was found to be both comfortable and accurate for continuous lung volume measurement.
机译:如果没有像面具这样的高度引人注目的仪器,目前尚无批准的临床方法可用于监测肺流量和容量。阻抗呼吸描记法(IP)似乎是建立准确的Holter型呼吸测量值的最有前途的方法。我们在静态条件下,使用带有纺织银线的胸带,从六种不同的四极电极配置中,测量了来自14位受试者的IP和肺活量计的体积信号。我们将肺活量计和IP的连续波形与Pearson线性相关系数p进行了比较,发现六种配置之间存在明显差异。对于最佳配置,对于所有受试者,p> 0.98时阻抗和肺活量计信号之间的差异几乎没有意义。在这些静态条件下,发现IP方法对于连续的肺体积测量既舒适又准确。