首页> 外文会议>World congress on medical physics and biomedical engineering;International congress of the IUPESM >Evaluation of an Integrated Tool for Automated Visualization of SubduralElectrodes
【24h】

Evaluation of an Integrated Tool for Automated Visualization of SubduralElectrodes

机译:硬膜下自动可视化集成工具的评估电极

获取原文

摘要

Using a 3D-demonstrator system for application in neurosurgery, an integrated tool for epilepsy surgery evaluation has been developed to visualize subdural electrodes derived from postoperative CT datasets on the cortical surface by 3D-rendering preoperative MRI datasets. Difficulties result from postoperative deformations caused by hematoma. To investigate deformation and resultant electrode localization errors, patient data, online database and also phantoms as a kind of well-known reference are used. A head phantom was designed to evaluate occurring errors, respectively. The developed tool works fully automated and takes about 2 minutes to visualize electrodes' position. The measured electrodes' position error does not fulfill clinical requirements yet; phantom data must be used for further investigations.
机译:使用3D演示器系统在神经外科中的应用,开发了一种用于癫痫手术评估的集成工具,以通过3D渲染术前MRI数据集可视化来自皮层表面术后CT数据集的硬膜下电极。困难是由血肿引起的术后变形引起的。为了研究变形和由此产生的电极定位误差,使用了患者数据,在线数据库以及体模作为一种众所周知的参考。头部模型被设计为分别评估发生的错误。开发的工具可以完全自动运行,大约需要2分钟才能看到电极的位置。测得的电极位置误差尚未达到临床要求;幻像数据必须用于进一步调查。