This paper presents the direct comparison between measurements and Monte Carlo ion chamber simulations employing the EGSnrc Monte Carlo codeSurrounding metal caps for a 1 cm3 ionization chamber (PTW30015) were varied both, in measurements and simulations. Absorbing caps made of iron, copper and aluminum alloy were manufactured each with wall thickness between 2 and 30 mm. As a radiation source, a Siemens Gammatron S ~(60)Co machine was used. Normalized dose values were compared and changes to the cross-sections used in the Monte Carlo simulations investigated.Results agreed within the range of the measurement uncertainties of -0.5%. Still, variations in photon cross-sections lead to significant changes in the results, highlighting the need of accurate input data for this type of investigation.
机译:本文介绍了使用EGSnrc蒙特卡洛代码进行的测量与蒙特卡洛离子室模拟之间的直接比较 在测量和模拟中,都改变了用于1 cm3电离室(PTW30015)的周围金属盖。制造了由铁,铜和铝合金制成的吸收帽,每个吸收帽的壁厚在2到30毫米之间。作为辐射源,使用了西门子Gammatron S〜(60)Co机器。比较归一化剂量值,并研究在蒙特卡洛模拟中使用的横截面变化。 结果在测量不确定度-0.5%的范围内一致。尽管如此,光子横截面的变化仍会导致结果发生重大变化,这突显了此类研究需要准确的输入数据。