Dose Distribution in Pediatric CT Abdominal Examination: Phantom Study

机译:小儿CT腹部检查的剂量分布:幻影研究

获取原文

摘要

To keep radiation doses during computed tomography (CT) examinations as low as reasonably achievable, performing a detailed dose measurement is important. A flexible acrylic sheet roll CT dosimetry phantom (SRCT-P) with radiochromic film (RF) was developed to estimate in detail the dose distribution during pediatric CT examination. The SRCT-Ps were elliptically-shaped by rolling up flexible acrylic sheets (1.1 g/cm'). The dose distributions in the SRCT-P (body thickness and width: 6-8 cm [neonates], 10-12 cm |infants], and 14-16 cm |three-year-o!d children|) were evaluated. RFs were positioned from the center to the surface along the long- and short-axis directions in each SRCT-P. The scanning parameters of the single detector CT were 120 kV, 250 mA, 1.0 sec/rot, a slice thickness of 5 mm, and a 1.0 beam pitch. When the mean center dose at 10-12 cm on the SRCT-P was taken as 100%, the mean center doses at 6-8 or 14-16 cm were 145 or 43%, respectively, and the mean surface doses for 0, 90, and 180 degrees at 6-8,10-12, and 14-16 cm of the SRCT-P were as follows: 158%, 159%, and 140%; 132%, 125%, and 116%; and 164%, 128%, and 131%, respectively. The center dose was increased with a decrease in the phantom size. The surface doses were decreased with decreasing phantom sizes of 14-16 to 10-12 cm; however, surface doses were increased with decreasing phantom sizes of 10-12 to 6-8 cm. The detail dose distribution of a CT examination can be measured separately by using a SRCT-P with a RF.
机译:为了使计算机断层扫描(CT)检查期间的放射剂量保持在合理可达到的最低水平,执行详细的剂量测量非常重要。开发了带有放射致变色膜(RF)的柔性丙烯酸薄板卷CT剂量测定体模(SRCT-P),以详细估计儿科CT检查期间的剂量分布。 SRCT-Ps通过卷起柔性丙烯酸板(1.1 g / cm')形成椭圆形。评估了SRCT-P中的剂量分布(身体厚度和宽度:6-8厘米(新生儿),10-12厘米|婴儿)和14-16厘米| 3岁儿童|)。在每个SRCT-P中,RF沿着长轴和短轴方向从中心到表面定位。单个检测器CT的扫描参数为120 kV,250 mA,1.0秒/转,5 mm的切片厚度和1.0的束间距。以SRCT-P上10-12厘米的平均中心剂量为100%时,6-8或14-16厘米的平均中心剂量分别为145或43%,平均表面剂量为0, SRCT-P在6-8、10-12和14-16厘米处分别为90度和180度,分别为:158%,159%和140%; 132%,125%和116%;和164%,128%和131%。中心剂量随着体模尺寸的减小而增加。随着体模尺寸从14-16到10-12 cm的减小,表面剂量减少了。然而,表面剂量随着幻影尺寸的减小(10-12至6-8 cm)而增加。可以使用带有RF的SRCT-P单独测量CT检查的详细剂量分布。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号