Energy Dependence Investigations of GafChromic EBT Films for Pretreatment Dosimetric Verification of IMRT Plans

机译:GafChromic EBT膜的能量依赖性研究,用于IMRT计划的预处理剂量学验证

获取原文

摘要

The purpose of the study was examination of energy dependence of Gafchromic EBT - a new radiochromic dosimetry film. The film response was measured irradiating the samples with doses from 0.1 Gy to 12 Gy using photon beams from the energy range 1.25 MV-25 MV. The samples were scanned and the film response was compared for different beam energies. The measurements were performed in order to evaluate potential use of radiochromic films in pretreatment dosimetric verifications of intensity modulated radiotherapy (IMRT) plans. A high uncertainty in readout of the film response was observed for films irradiated with doses lower than 1 Gy. The relative difference exceeds 20 % for doses below 1 Gy while for doses over 1 Gy the measured film response differs by less than 5 % for the whole examined energy range. No energy dependence of Gafchromic EBT film response was observed in the investigated energy range but the doses measured for pretreatment plan verifications should exceed 1 Gy.
机译:该研究的目的是检查Gafchromic EBT的能量依赖性-一种新的放射致变色剂量测定膜。使用能量范围为1.25 MV-25 MV的光子束,以0.1 Gy至12 Gy的剂量照射样品,从而测量薄膜响应。扫描样品,并比较不同光束能量的胶片响应。进行测量是为了评估放射性变色膜在强度调制放射治疗(IMRT)计划的预处理剂量学验证中的潜在用途。对于剂量低于1 Gy的胶片,观察到的胶片响应读数存在很高的不确定性。对于小于1 Gy的剂量,相对差异超过20%,而对于大于1 Gy的剂量,在整个检查的能量范围内,测得的薄膜响应差异小于5%。在研究的能量范围内未观察到Gafchromic EBT膜响应的能量依赖性,但用于预处理计划验证的剂量应超过1 Gy。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号