Axial and transverse displacement tolerances during excimer laser surgery for myopia

机译:近视激光手术期间的轴向和横向性耐受性

获取原文

摘要

This paper presents an analysis of the effects of axial and transverse displacement on the optical quality and accuracy of lenses created during excimer laser photoablation. Tolerance levels for axial positioning of the cornea prior to and during surgery are presented. The axial tolerance levels are dependent upon a number of parameters which include the intended dioptric correction and laser system cone angle. A collimation lens is introduced as a means of desensitizing the laser system to axial displacement. Transverse displacement tolerances during laser treatment are shown to depend upon the treatment diameter, dioptric correction and acceptable distortion level in the lens ablated into the anterior corneal stroma. A video and computer analysis of transverse motion during seven randomly selected excimer laser refractive surgeries is presented. Although transverse displacement exceeded the tolerance levels presented, it did not appear to affect the quality of correction in the eight patients analyzed.
机译:本文介绍了轴向和横向位移对准分激光光电激光拍摄期间产生的光学质量和精度的影响的分析。介绍了在手术前和手术期间的角膜轴向定位的耐受性水平。轴向公差水平取决于许多包括预期屈光度校正和激光系统锥角的参数。将准直透镜作为脱敏激光系统以轴向位移的手段。激光处理期间的横向位移耐受性被显示为取决于镜片中的透镜中的治疗直径,屈光度校正和可接受的变形水平烧蚀到前角膜基质中。介绍了七种随机选择的准分子激光屈光屈光手术过程中横向运动的视频和计算机分析。虽然横向位移超过呈现的耐受性水平,但它没有似乎影响八名患者分析的矫正质量。