Temperature increase during in vitro 308 nm excimer laser ablation of porcine aortic tissue

机译:猪主动脉组织体外308 nm受激准分子激光消融过程中的温度升高

获取原文

摘要

Abstract: The temperature increase of porcine aortic tissue irradiated with 220 ns XeCl 308 nm excimer laser pulses above ablation threshold have been measured by means of an IR camera. The decay of the temperature increase after a laser pulse is characterized by a fast decay during the first few milliseconds followed by a slow decay over a prolonged period, suggesting temperature accumulation at repetition rates used in laser angioplasty. The decay is adequately described by the theoretical prediction of the diffusion of heat after adiabatic heating of the tissue by a short laser pulse. From the temporal behavior of the decrease of temperature after a laser pulse, attenuation coefficients at 308 nm could be estimated. Attenuation coefficients after a single pulse are smaller than attenuation coefficient after multiple pulses. The temperature increases found in the experiments might explain thermal damage in myocardial tissue in earlier experiments. In an experimental set up simulating the situation during clinical excimer laser angioplasty, temperature increases of 43$DGR@C average were found.!
机译:摘要:用红外热像仪测量了高于消融阈值的220 ns XeCl 308 nm受激准分子激光脉冲辐照的猪主动脉组织的温度升高。激光脉冲后温度升高的衰减特征是在最初的几毫秒内快速衰减,然后在很长一段时间内缓慢衰减,这表明温度在激光血管成形术中以重复速率积累。衰减是通过短激光脉冲对组织进行绝热加热后的热扩散的理论预测来充分描述的。根据激光脉冲后温度下降的时间行为,可以估算出308 nm处的衰减系数。单个脉冲后的衰减系数小于多个脉冲后的衰减系数。实验中发现的温度升高可能可以解释早期实验中心肌组织的热损伤。在模拟临床受激准分子激光血管成形术过程中的情况的实验装置中,发现温度平均升高了43 $ DGR @ C。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号