首页> 外文会议> >Body-Contour Acquisition Versus Circular Orbit Acquisition with Resolution Recovery in Cardiac SPECT
【24h】

Body-Contour Acquisition Versus Circular Orbit Acquisition with Resolution Recovery in Cardiac SPECT

机译:身体轮廓获取与圆形轨道获取以及心脏SPECT中的分辨率恢复

获取原文

摘要

Two approaches are often applied to reduce the effects of depth-dependent resolution in SPECT. First is to use body-contour acquisition orbits and second is to apply resolution recovery (RR) in reconstruction. Unfortunately it is sometimes difficult to combine these methods, because in body-contour acquisitions especially many older gamma cameras do not record the object-to-detector distances, which are needed to perform RR in reconstruction. The objective of this study is to investigate if it would is better to use body-contour orbit without RR or circular orbit with RR in cardiac SPECT. The results of simulation and physical phantom studies showed that circular orbit with RR produce images with lower bias, higher contrast and better resolution than the body-contour orbit without RR. Therefore if object-to-detector distances are not available for body-contour orbits circular orbits with RR should be preferred in cardiac SPECT.
机译:通常使用两种方法来减少SPECT中依赖深度的分辨率的影响。首先是使用体轮廓获取轨道,其次是在重建中应用分辨率恢复(RR)。不幸的是,有时很难组合这些方法,因为在人体轮廓采集中,尤其是许多较旧的伽玛相机无法记录物体到检测器的距离,而距离是重建中执行RR所必需的。这项研究的目的是研究在心脏SPECT中使用无RR的体轮廓轨道或有RR的圆形轨道是否更好。模拟和物理幻像研究的结果表明,与没有RR的人体轮廓轨道相比,具有RR的圆形轨道产生的图像具有更低的偏差,更高的对比度和更好的分辨率。因此,如果身体轮廓轨道无法达到检测器到检测器的距离,则在心脏SPECT中应优先选择带RR的圆形轨道。