Ultrasonic measurement of flow-mediated vasodilatation: clinical results

机译:超声测量血流介导的血管舒张:临床结果

获取原文

摘要

With respect to brachial artery flow-mediated vasodilatation (FMD), which is frequently used as a measure of endothelial function, it is necessary to observe diameter changes of small arterial vessels non-invasively for several minutes at a high resolution. We developed a system composed of a Siemens Omnia ultrasound system with a VF13-5 transducer (9 MHz center frequency) and a personal computer, that is capable of calculating vessel diameter changes with an accuracy below the wavelength of the ultrasound system in real-time at a frame rate of 27 Hz. We could verify experimentally that the accuracy of the measurements is below 1 /spl mu/m. Results from clinical measurements of carotid and brachial arteries show that arterial pulsations below 0.1 mm can be visualized reliably over several minutes. Furthermore, we can present examples from clinical FMD measurements, where the pulsation of the brachial artery is visible as well as an increase in arterial diameter after the deflation of a forearm cuff, which was used to produce an ischaemia for 5 minutes.
机译:对于经常被用作测量内皮功能的肱动脉血流介导的血管舒张(FMD),有必要以高分辨率无创地观察小动脉血管的直径变化,持续数分钟。我们开发了一个系统,该系统由带有VF13-5传感器(中心频率为9 MHz)的Siemens Omnia超声系统和个人计算机组成,该系统能够实时以低于超声系统波长的精度计算血管直径变化以27 Hz的帧频我们可以通过实验验证测量的精度低于1 / spl mu / m。颈动脉和肱动脉的临床测量结果表明,在几分钟内可以可靠地看到0.1 mm以下的动脉搏动。此外,我们可以提供来自临床FMD测量的示例,其中前臂袖带放气后可见肱臂的搏动以及动脉直径的增加,前臂袖带用于产生5分钟的局部缺血。