首页> 外文会议> >Calibration of A Low-Temperature Cable-Less Lightpipe Pyrometer on The NIST Post-Exposure Bake Test Bed
【24h】

Calibration of A Low-Temperature Cable-Less Lightpipe Pyrometer on The NIST Post-Exposure Bake Test Bed

机译:在NIST曝光后烘烤测试台上校准低温无电缆光导管高温计

获取原文
获取外文期刊封面目录资料

摘要

The advent of the cable-less lightpipe radiation thermometer (CLRT) has resulted in a significant improvement in the accuracy of lightpipe radiation thermometer calibrations and measurements. CLRT systems show great promise in noncontact measurements by the elimination of the uncertainties caused by the long fiber optic cables and their connections and by the extension of the spectral range to handle low-temperature applications down to room temperature. A CLRT was first calibrated with the oil-bath and water-bath blackbody sources from 40°C to 180°C. Then the CLRT was compared to thin-film thermocouples and platinum resistance thermometers on a silicon wafer heated in a Post-Exposure Bake (PEB) Test Bed. Comparison of the CLRT with both the blackbody and thermocouple standards provides confidence in using CLRTs and allows researchers to continue research into improving the accuracy and feasibility of applying CLRTs in semiconductor processing.
机译:无电缆光导管辐射温度计(CLRT)的出现极大地提高了光导管辐射温度计的校准和测量精度。 CLRT系统在非接触式测量中显示出巨大的希望,因为它消除了由长光纤电缆及其连接所引起的不确定性,并扩展了光谱范围以应对低温至室温的应用。首先使用40°C至180°C的油浴和水浴黑体源对CLRT进行校准。然后将CLRT与在曝光后烘烤(PEB)测试床上加热的硅晶片上的薄膜热电偶和铂电阻温度计进行比较。将CLRT与黑体标准和热电偶标准进行比较可为使用CLRT提供信心,并使研究人员能够继续进行研究,以提高将CLRT用于半导体加工的准确性和可行性。